A 3-DIMENSIONAL ANALYSIS OF SLOTTED TUBE RESONATOR FOR MRI

Citation
Qa. Chen et al., A 3-DIMENSIONAL ANALYSIS OF SLOTTED TUBE RESONATOR FOR MRI, IEEE transactions on medical imaging, 13(4), 1994, pp. 587-593
Citations number
16
Categorie Soggetti
Engineering, Biomedical","Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
02780062
Volume
13
Issue
4
Year of publication
1994
Pages
587 - 593
Database
ISI
SICI code
0278-0062(1994)13:4<587:A3AOST>2.0.ZU;2-U
Abstract
A three dimensional model of a slotted tube resonator (STR) used as a probe in the magnetic resonance imaging (MRI), which is loaded by a di electric body and surrounded by a conducting shield, is analyzed by us ing the variational method and the dyadic Green's function of a circul ar waveguide having a dielectric core. Three surface current modes are properly assumed to expand the currents on the STR. The characteristi cs such as the input impedance, the resonance frequency, the Q value, and the magnetic field distribution are obtained to show the effects o f the dielectric body and the conducting shield. Some theoretical resu lts are compared with the measured data to confirm the validity of the present analysis.