PATIENT POSITIONING FOR FRACTIONATED PRECISION RADIATION TREATMENT OFTARGETS IN THE HEAD USING FIDUCIAL MARKERS

Citation
E. Grusell et al., PATIENT POSITIONING FOR FRACTIONATED PRECISION RADIATION TREATMENT OFTARGETS IN THE HEAD USING FIDUCIAL MARKERS, Radiotherapy and oncology, 33(1), 1994, pp. 68-72
Citations number
8
Categorie Soggetti
Oncology,"Radiology,Nuclear Medicine & Medical Imaging
Journal title
ISSN journal
01678140
Volume
33
Issue
1
Year of publication
1994
Pages
68 - 72
Database
ISI
SICI code
0167-8140(1994)33:1<68:PPFFPR>2.0.ZU;2-5
Abstract
When irradiating targets in the brain, an accurately localised dose is often needed. One crucial moment to achieve this is the positioning o f the patient. We have developed a positioning method where the patien t is immobilised with a bite block and a head mould, and where the pos ition of the target is determined by X-ray imaging of fiducial markers that are placed in the patient's skull. A method for computing the tr ansformation needed to move the target from the observed to the prescr ibed position and orientation is described. This method uses the infor mation from two orthogonal X-ray images and takes measurement errors a nd data from three or more markers into account. Results from using th e method clinically in proton beam therapy are given.