R. Mischke et D. Menzel, MEASURING FIBRINOGEN CONCENTRATIONS IN HE ALTHY DOGS - STANDARDIZATION, COMPARISON OF METHODS AND REFERENCE VALUES, Journal of veterinary medicine. Series A, 41(8), 1994, pp. 587-598
Plasma fibrinogen concentration was measured in 67 healthy, adult dogs
using five different methods (gravimetry, methods described by JACOBS
SON (1955) RATNOFF and MENZIE (1951), and CLAUSS (1957), and functiona
l photometric assay). Apart from using linear regression and the Pears
on correlation coefficient (r) in order to characterize the relation b
etween different methods, reference ranges (2.5-97.5%-fractile) were c
alculated for all methods. For calibration of the CLAUSS method (1957)
and the photometric assay, dog plasma with a defined fibrinogen conce
ntration was used. Measurements of commercial human fibrinogen standar
ds yielded a good conformity with the concentrations specified by the
manufacturer (values approximately 3% too low). These standards appear
, therefore, to be also suited to the calibration of measurements of d
og fibrinogen. The reference range for the gravimetry was 1.08-2.88 g
fibrinogen per litre of plasma. A considerable conformity and close co
rrelation was seen between the fibrinogen concentration measured by gr
avimetry and by using methods described by JACOBSSON (1955; y = 1.088
x -0.142, r = 0.967) or CLAUSS (1957; y = 0.999 x -0.004, r = 0.973),
respectively. Between the reference-method gravimetry and the photomet
ric method and RATNOFF-MENZIE (1951) method, respectively, a less clos
e correlation, as well as a minor conformity, was found.