THE ACCURACY OF CREATININE METHODS BASED ON THE JAFFE REACTION - A QUESTIONABLE MATTER

Citation
Bg. Blijenberg et Hj. Brouwer, THE ACCURACY OF CREATININE METHODS BASED ON THE JAFFE REACTION - A QUESTIONABLE MATTER, European journal of clinical chemistry and clinical biochemistry, 32(12), 1994, pp. 909-913
Citations number
18
Categorie Soggetti
Biology,"Chemistry Medicinal
ISSN journal
09394974
Volume
32
Issue
12
Year of publication
1994
Pages
909 - 913
Database
ISI
SICI code
0939-4974(1994)32:12<909:TAOCMB>2.0.ZU;2-Z
Abstract
The determination of creatinine in serum based on the Jaffe reaction w as evaluated with four current analysers. In particular, the comparabi lity of results was determined also with survey specimens. Recalibrati on of 3 out of 4 modifications was necessary, based on the results of patient samples as verified with a HPLC-method. One of the methods pro ved to give an unacceptable scatter for the results in the lower range (30-150 mu mol/l). A limited interference study (haemoglobin, lipids, bilirubin and acetone) and a method assessment with quality control s era supported the conclusion that the overall accuracy of creatinine m ethods based on the Jaffe reaction is questionable.