Wl. Straube et al., PHASE-STABILITY OF A CLINICAL PHASED-ARRAY SYSTEM FOR DEEP REGIONAL HYPERTHERMIA, International journal of hyperthermia, 11(1), 1995, pp. 87-93
Measurements were performed on the amplifiers of a phased array system
to evaluate the dependence of phase on frequency. The power from the
amplifiers was terminated into a 50 ohm load at the point of connectio
n to the antennas of the applicator, and the power was sampled to dete
rmine the phase relation between amplifier channels. It was found that
the measured phase difference between two amplifier channels can: (1)
change by as much as 20 degrees with frequency; (2) be as much as 12
degrees different than the prescribed phase; and (3) be as much as 30
degrees different than the phase measured and displayed by the system.
Previous studies indicate that in order to accurately plan and delive
r treatments with this type of device, it is necessary to know the pha
se relationship of the array to within at least 20 degrees. Since diff
erences as great as 20 degrees were obtained under ideal loading condi
tions, greater differences could result during normal (clinical) loadi
ng conditions, and these may not be compensated by the operator becaus
e of the inaccurate values reported by the system itself. These findin
gs should be of concern to investigators using these devices since the
y could result in SAR distributions different from those planned and/o
r optimized for a particular patient. It is therefore recommended that
the measurements such as those reported here also be performed on sim
ilar clinical devices as a standard quality assurance procedure so tha
t the power steering capabilities of these systems can be utilized eff
ectively and safely.