Tb. Watson, EVALUATION OF AN INTENSIVE SAMPLING AND ANALYSIS METHOD FOR CARBON-MONOXIDE, Journal of the Air & Waste Management Association [1995], 45(1), 1995, pp. 29-35
This paper describes a portable, low-cost whole air sampler capable of
collecting 12 1-L samples in tedlar bags and using a single ''D'' cel
l for power. The results are presented from two tests of an intensive
sampling and analysis method for measuring ambient carbon monoxide (CO
) concentrations in urban areas using this sampler. The method is eval
uated in comparison with an approved continuous CO analyzer, in interc
omparisons of the results of re-analysis of the samples, and by examin
ation of the results from co-located samplers. The precision of the an
alytical method was found to be +/-0.30 ppmv. The precision of the sam
pling method was found to be +/-0.73 ppmv and therefore is the limitin
g factor in the method's overall precision. These values are sufficien
t to verify attainment of National Ambient Air Quality Standards (NAAQ
S) levels in urban areas. Improvements in the sampler and analytical p
rocedure are discussed.