LABORATORY TESTING OF 3 NEW ANESTHESIA VE NTILATORS

Authors
Citation
L. Beydon et N. Liu, LABORATORY TESTING OF 3 NEW ANESTHESIA VE NTILATORS, Annales francaises d'anesthesie et de reanimation, 13(6), 1994, pp. 807-810
Citations number
NO
Categorie Soggetti
Anesthesiology
ISSN journal
07507658
Volume
13
Issue
6
Year of publication
1994
Pages
807 - 810
Database
ISI
SICI code
0750-7658(1994)13:6<807:LTO3NA>2.0.ZU;2-F
Abstract
We report the bench testing of three new anaesthesia ventilators : Fle xima(TM) (Datex), SA2/RA2(TM) and Cato(TM) (Drager). The test circuit included a two compartment lung model, a pneumotachograph and a pressu re gauge. Volume was integrated from flow signal. The tidal volume del ivered by the Flexima(TM) and SA2/RA2(TM) ventilators decreased with i ncreasing mechanical work load contrarily to the Cato(TM). Moreover, t he tidal volume of the Flexima(TM) increased with the fresh gas flow. In spontaneous ventilation inspiratory resistance were low.