CLASSIFICATION OF ESTRADURIN(R) BATCHES - CORRELATION BETWEEN P-31 NMR AND A BIOLOGICAL DURATION TEST FOR BATCH APPROVAL

Citation
B. Thelin et al., CLASSIFICATION OF ESTRADURIN(R) BATCHES - CORRELATION BETWEEN P-31 NMR AND A BIOLOGICAL DURATION TEST FOR BATCH APPROVAL, Chemometrics and intelligent laboratory systems, 27(1), 1995, pp. 135-145
Citations number
7
Categorie Soggetti
Computer Application, Chemistry & Engineering","Instument & Instrumentation","Chemistry Analytical","Computer Science Artificial Intelligence","Robotics & Automatic Control
ISSN journal
01697439
Volume
27
Issue
1
Year of publication
1995
Pages
135 - 145
Database
ISI
SICI code
0169-7439(1995)27:1<135:COEB-C>2.0.ZU;2-Z
Abstract
The analyses used for batch approval of Estradurin(R) are a biological duration test on mice together with gel permeation chromatography. Fo r 38 historical batches these analyses have been complemented with P-3 1 NMR analysis. In this work we have used multivariate methods such as principal components analysis, discriminant projections to latent str uctures and pattern recognition to find a correlation between the old and the new analyses for batch approval. In this way it has been possi ble to establish a mathematical model for approving/not approving batc hes of Estradurin(R) based on 31 P NMR analysis. This should make it p ossible to replace the animal duration test.