THERMOLUMINESCENCE DOSIMETER FOR PERSONAL DOSE-EQUIVALENT ASSESSMENT

Citation
Ta. Dasilva et al., THERMOLUMINESCENCE DOSIMETER FOR PERSONAL DOSE-EQUIVALENT ASSESSMENT, Radiation protection dosimetry, 58(1), 1995, pp. 17-21
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
ISSN journal
01448420
Volume
58
Issue
1
Year of publication
1995
Pages
17 - 21
Database
ISI
SICI code
0144-8420(1995)58:1<17:TDFPDA>2.0.ZU;2-X
Abstract
The possiblity was investigated of utilising a Brazilian thermolumines cence individual dosemeter, usually calibrated in terms of photon dose equivalent, for the assessment of the personal dose equivalent, H(p)( d), at depths of 0.07 and 10 mm. The dosemeter uses four CaSO4:Dy ther moluminescent detectors, between different filters, as the sensitive m aterials. It was calibrated in gamma and X radiation fields in the ene rgy range from 17 to 1250 keV. Linear combinations of the responses of three detectors, in this energy range, allow the evaluation of H(p)(0 .07) and H(p)(10), for radiation incidence angles varying from 0 to 60 degrees, with an accuracy better than 35%. The method is not applicab le to mixed photon-beta fields.