A. Kojlo et Jm. Calatayud, FIA-FLUOROMETRIC DETERMINATION OF ADRENALINE BY OXIDATION WITH A SOLID-PHASE REACTOR OF MANGANESE-DIOXIDE INCORPORATED IN POLYESTER RESIN BEADS, Analytical letters, 28(2), 1995, pp. 239-247
The FIA-spectrofluorimetric determination of adrenaline was carried ou
t by reaction of the drug with manganese dioxide entrapped in a polyme
ric material in a solid-chase reactor the oxidized drug was monitored
fluorimetrically at 540 nm (I-oxo 330.0 nm). The calibration graph for
adrenaline was linear over the range 0.5 - 20 mu g ml(-1) with a rela
tive standard deviation of 2.0% (at 5 ug ml(-1)) and the sample throug
hout of 65 h(-1). The influence of foreign compounds was studied and t
he method was applied to the determination of adrenaline content in a
pharmaceutical formulation.