R. Nicoletti et al., SPET QUALITY-CONTROL - A COMPREHENSIVE TEST PROCEDURE FOR EVALUATING THE ROTATION SYSTEM PARAMETERS COR-OFFSET, Y-SHIFT AND DETECTOR TILT, Nuclear medicine communications, 16(2), 1995, pp. 105-111
This paper describes a method for SPET quality control that facilitate
s the determination of COR-offset, parallel y-shift and detector tilt
by means of a single, comprehensive test using a special phantom consi
sting of three point sources. The evaluation of the method was perform
ed by a series of measurements utilizing manually adjusted detector ti
lts and a simulation of y-shift by moving the phantom. The results sho
w that y-shift and detector tilt are measured with reasonable accuracy
, and that y-shift caused by variations in detector tilt can be differ
entiated from y-shift caused by other factors.