R. Torralba et al., FLOW-INJECTION ANALYSIS AND CONTINUOUS-FLOW SYSTEMS TO DETERMINE AS(III) AND AS(V) BY HYDRIDE GENERATION ATOMIC-ABSORPTION SPECTROMETRY, Analusis, 22(10), 1994, pp. 478-482
A method for the determination of total arsenic and inorganic As(III)
and As(V) by either FIA or continuous flow hydride generation-AAS is r
eported. If AsH3 is generated in 6 M HCl, the addition of KI to reduce
As(V) to As(III) is not required. The precision of arsine generation
in 6 M HCl or citric/citrate medium is higher in the continuous flow s
ystem than in the FIA system. The detection limits in 6 M HCl medium a
re reported for 0.75 ng of As(III) in FIA and for 4.5 ng ml(-1) of As(
III) in continuous flow. The proposed method has been applied to drink
ing waters and the total arsenic determination has been validated for
water analysis by a laboratory intercomparison study.