A precisely working automated system for the investigation of phosphol
ipases D (PLDs, EC 3.1.4.4) from plant and microbial sources with how
injection analysis (FIA) has been developed. The two versions of the F
IA setup described are based on the oxidation of choline liberated fro
m phosphatidylcholine by PLD action and catalyzed by choline oxidase a
nd the chemiluminescence detection of hydrogen peroxide produced by th
is reaction. The correlation between this chemiluminescence signal and
the PLD activity was linear in the range between 1 and 100 mU/ml PLD.
The sampling frequency was 12 samples per hour. This method was used
to compare three different PLDs from cabbage and microbial sources wit
h respect to their pH optima, temperature stability, effecters, and v/
[S] characteristics. (C) 1997 Academic Press, Inc.