Aj. Compton et al., ARRHYTHMIA RECOGNITION STRATEGIES AND HARDWARE DECISIONS FOR THE IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - A REVIEW, Medical engineering & physics, 17(2), 1995, pp. 96-103
The avoidance of inappropriate shocks from the implantable cardioverte
r-defibrillator (ICD), together with its need to apply antiachycardia
pacing to either atria or ventricles, demands considerable sophisticat
ion in the design of algorithms to interpret electrical or other cardi
ac signals in real-time, Methods based on rate and using single short-
gap bipolar leads lack discrimination. Right ventricular electrogram m
orphology algorithms offer improvement but no universal algorithm exis
ts; however, for any given patient an optimum algorithm of this type m
ight be found. One improvement would be to provide atrial information
in addition, by employing more than one electrode or a long-gap single
bipolar lead. Alternatives transducer signals could be included, once
their efficacy and reliability have been improved. A different approa
ch would be to use the much more sophisticated algorithms at present b
eing tried with surface electrocardiograms. Integrated Circuit technol
ogy is reaching the point where this could be done but the requirement
for exceptionally high reliability means that special system structur
es, such as a Memory intensive Computer Architecture, may be required.
When decisions on these approaches are to be made, it must also be re
membered that ICDs will soon be implanted and programmed as a routine
rather than highly specialized procedure.