Background. The Hemopump (DLP/Medtronic) has been in clinical we for a
bout 7 years. There is still no adequate way of determining actual out
put from the three available pump systems in the clinical situation. I
f the pump is completely stopped during weaning from the device, there
is a possibility of back-leakage through the pump, endangering the pa
tient from regurgitation into the left ventricle. It can also make it
more difficult to judge the recovery of heart function because of a vo
lume load of the left ventricle. The aim of this study was to evaluate
in a standardized, experimental in vitro model the output from three
different-sized Hemopump catheters at various pressure levels and to q
uantify the backflow through the pumps. Methods. The Hemopump models w
ere tested in an in vitro study regarding total outflow at various spe
eds at three pressure levels. The back-how through the pumps was also
measured with the pumps at a complete stop. Results. The outflow from
the Hemopumps ranged from 0.4 to 4.5 L/min, depending on which pump an
d speed were used. Variations in total output, depending on speed and
various pressure settings, could be up to 0.4 L/min. Back-flow through
the pump into the left ventricle may he as great as 1.6 L/min. Conclu
sions. The flow outputs from the different Hemopump models were reprod
ucible over time and were closely related to the resistance of the mod
el. The Hemopump, if not running can induce substantial regurgitation
through the pump into the left ventricle. (C) 1997 by The Society of T
horacic Surgeons