NEW STANDARDLESS METHOD OF PB ANALYSIS BY ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROSCOPY IN AIR AEROSOLS - COMPARISON WITH THE OFFICIAL METHOD

Citation
C. Locatelli et al., NEW STANDARDLESS METHOD OF PB ANALYSIS BY ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROSCOPY IN AIR AEROSOLS - COMPARISON WITH THE OFFICIAL METHOD, Applied spectroscopy, 50(12), 1996, pp. 1585-1589
Citations number
8
Categorie Soggetti
Instument & Instrumentation",Spectroscopy
Journal title
ISSN journal
00037028
Volume
50
Issue
12
Year of publication
1996
Pages
1585 - 1589
Database
ISI
SICI code
0003-7028(1996)50:12<1585:NSMOPA>2.0.ZU;2-J
Abstract
A standardless analysis method with electrothermal atomization/ atomic absorption spectroscopy (ETA-AAS) measurements is proposed and applie d to the determination of Pb in an urban environment (Bologna, Italy) and compared with results obtained in the same place and time, with th e method dictated by Italian law, which is derived by European Communi ty Directives, It was found that the difference previously observed be tween the two methods is due to the low efficiency of the 0.45-mu m-po re-size paper filters, dictated by the official method, The uncovering of the error source validates the proposed method and highlights its accuracy, Moreover, the two methods can be conveniently combined to gi ve two series of data relevant to pollutants associated with low- and high-dimension particles, This capability gives a much more precise pi cture of the hazard of a pollutant.