At cardiac investigations Doppler echocardiography is an established t
echnique for the recording of intracardial and intravascular now veloc
ities. Transvalvular and venous flows are, however, markedly influence
d by respiration. Since the start of inspiration is the important time
of reference, accurate recording of the respiratory phase is importan
t when analysing these flow velocities. A bioacoustic technique was th
erefore adopted to meet this demand and was tested in 10 normal subjec
ts. Oesophageal balloon technique was used as a reference, and the bio
acoustic technique was compared with a respiratory inductive plethysmo
graph. The average delay for the acoustic sensor signal compared to th
e oesophageal pressure changes was 205+/-46 (SD) ms, ranging from 160
to 320 ms for normal breathing, and is comparable to that of the respi
ratory inductive plethysmograph (210+/-90 ms). The bioacoustic techniq
ue is easier to apply than the respiratory inductive plethysmograph. T
he sensitivity to disturbances can be further reduced by signal proces
sing and the method has clinical potential for the future.