IMAGE QUALITY AND DOSE DISTRIBUTION IN MULTISCAN AND FAST RING CT SYSTEMS

Citation
F. Levrero et al., IMAGE QUALITY AND DOSE DISTRIBUTION IN MULTISCAN AND FAST RING CT SYSTEMS, Radiation protection dosimetry, 57(1-4), 1995, pp. 135-138
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
ISSN journal
01448420
Volume
57
Issue
1-4
Year of publication
1995
Pages
135 - 138
Database
ISI
SICI code
0144-8420(1995)57:1-4<135:IQADDI>2.0.ZU;2-5
Abstract
Modem CT systems allow examinations with both multiscan and volumetric techniques. This study on image quality and dose distribution was car ried out varying the operating conditions (step and slice thickness) f or each technique. To describe the physical image quality two basic pa rameters were taken into account: spatial resolution and noise. The me asurements were performed using an acrylic phantom, which includes a s eries of inserts for the response to spatial frequencies and a uniform insert for noise evaluation. The phantom images recorded on the film by a multiformat laser printer were acquired with a TV digitiser syste m and processed with a customised software package. Dosimetric evaluat ions were performed along the longitudinal axis using an air density e quivalent phantom holding the TL dosemeters at the desired location.