IMPACT OF SEVERAL RECOMMENDED ACTIONS FOR IMPROVING IMAGE QUALITY IN MAMMOGRAPHY

Citation
M. Chevalier et al., IMPACT OF SEVERAL RECOMMENDED ACTIONS FOR IMPROVING IMAGE QUALITY IN MAMMOGRAPHY, Radiation protection dosimetry, 57(1-4), 1995, pp. 155-158
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
ISSN journal
01448420
Volume
57
Issue
1-4
Year of publication
1995
Pages
155 - 158
Database
ISI
SICI code
0144-8420(1995)57:1-4<155:IOSRAF>2.0.ZU;2-V
Abstract
The dose/image quality relationship of mammography in the area of Madr id was studied during the period 1989-91. As a conclusion, several act ions were recommended to the facilities and its impact was evaluated d uring the period 1992-93. The results showed that the replacement of t he old X ray units with modem ones provided the major improvement in t he image quality. Conversely, the breast doses were increased due to t he presence of a grid (64%). However, this increase was balanced by th e major improvement found in image quality. The replacement of the ima ge receptors by high contrast screen-film combinations did not produce a noticeable variation in the image quality. However, this action was important in reducing the breast doses (62%). Overall the actions res ulting from quality control recommendations slightly improved the imag e quality but the doses were reduced by 41%.