M. Chevalier et al., IMPACT OF SEVERAL RECOMMENDED ACTIONS FOR IMPROVING IMAGE QUALITY IN MAMMOGRAPHY, Radiation protection dosimetry, 57(1-4), 1995, pp. 155-158
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
The dose/image quality relationship of mammography in the area of Madr
id was studied during the period 1989-91. As a conclusion, several act
ions were recommended to the facilities and its impact was evaluated d
uring the period 1992-93. The results showed that the replacement of t
he old X ray units with modem ones provided the major improvement in t
he image quality. Conversely, the breast doses were increased due to t
he presence of a grid (64%). However, this increase was balanced by th
e major improvement found in image quality. The replacement of the ima
ge receptors by high contrast screen-film combinations did not produce
a noticeable variation in the image quality. However, this action was
important in reducing the breast doses (62%). Overall the actions res
ulting from quality control recommendations slightly improved the imag
e quality but the doses were reduced by 41%.