Nj. Uys et al., A NONINVASIVE METHOD TO CONTROL THE TUBE CURRENT CALIBRATION OF DIAGNOSTIC-RADIOLOGY EQUIPMENT, Radiation protection dosimetry, 57(1-4), 1995, pp. 227-229
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
The correct calibration of the tube current of diagnostic X ray equipm
ent is important to ensure optimal image quality. This decreases the n
umber of retakes which will reduce the radiation dose to the patient a
nd the radiation worker. The direct measurement of the tube current is
time-consuming and is not part of the non-invasive quality control pr
ogramme. A technique was therefore developed where readings from a NER
O system were used to control the tube current calibration of diagnost
ic X ray equipment. If the tube potential and timer are accurately cal
ibrated, it is possible to derive the consistency of the tube current
calibration from the exposure measurement. The exposure of 27 X ray ma
chines was measured using a 65 cm focus-detector distance at 80 kV. Pr
evious tests on the different machines showed that the calibration of
the timer and tube potential was correct within 5%. The half value lay
er (HVL) for each machine was determined at 80 kV. The range of HVL va
lues was from 2.44 to 3.62 mm Al at 80 kV and the corresponding exposu
re from 0.06 to 0.19 mGy (mA.s)(-1) (95% confidence level). If the exp
osure is not within these limits with a correct tube potential and tim
er calibration, it will be indicative of a faulty tube current value.
A non-invasive method was developed to control the tube current calibr
ation of diagnostic X ray machines and this study showed that it could
be implemented successfully.