ANALYSIS OF THE RESULTS OF A QC PROJECT ON MAMMOGRAPHY IN GREECE

Citation
A. Flionivyza et al., ANALYSIS OF THE RESULTS OF A QC PROJECT ON MAMMOGRAPHY IN GREECE, Radiation protection dosimetry, 57(1-4), 1995, pp. 329-332
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
ISSN journal
01448420
Volume
57
Issue
1-4
Year of publication
1995
Pages
329 - 332
Database
ISI
SICI code
0144-8420(1995)57:1-4<329:AOTROA>2.0.ZU;2-5
Abstract
Quality Control (QC) has been carried out on twenty nine mammographic units installed in public hospitals, all over Greece. The QC procedure includes: registration of general information and technical descripti on of the mammographic chain, QC checks concerning environmental condi tions, radiation safety, beam characteristics, accessories and periphe rals performance, as well as image quality and patient dose. The colle cted data are characterised by great non-uniformity and in some cases indicate a poor performance level. Analysis of the results helps to im prove the existing situation with reference to patient dose and image quality.