A. Calzado et al., A COMPARISON OF MEASURED AND CALCULATED ORGAN DOSES FROM CT EXAMINATIONS, Radiation protection dosimetry, 57(1-4), 1995, pp. 381-385
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
Organ doses from a set of frequent CT examinations have been estimated
from measurements in a physical anthropomorphic phantom (Remab system
) by using thermoluminescence dosemeters. For the same examination tec
hniques, organ dose coefficients (taken from the literature) obtained
by Monte Carlo techniques and using mathematical phantoms. The results
arrived at by the two methods are compared, trying to explain the mos
t significant differences and their influence on the estimated values
of effective dose. The experimental and calculated outcomes from such
simulations are also compared to the mean dosimetric results on patien
ts from a 1991 regional survey of CT practice in the area of Madrid. S
ome comments about the complementary use of information coming from bo
th methods are made.