DOSE PROFILE AND DOSE INDEX ANALYSIS IN COMPUTED-TOMOGRAPHY

Citation
Ad. Oliveira et al., DOSE PROFILE AND DOSE INDEX ANALYSIS IN COMPUTED-TOMOGRAPHY, Radiation protection dosimetry, 57(1-4), 1995, pp. 387-391
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
ISSN journal
01448420
Volume
57
Issue
1-4
Year of publication
1995
Pages
387 - 391
Database
ISI
SICI code
0144-8420(1995)57:1-4<387:DPADIA>2.0.ZU;2-D
Abstract
A function to describe the dose profile in Computed Tomography (CT) be ams is presented. To measure the dose profile free-in-air a set of 20 thermoluminescence dosemeters was used. The symmetry or asymmetry, lef t and right penumbrae, length of the plateau and full width at half ma ximum (FWHM) are function parameters, determined with the fit to the T LD data, allowing a thorough characterisation of the beam. A new CTDI (designated CTDI) is defined using the FWHM instead of the nominal sl ice thickness, which is physically the correct normalising factor, pro ducing a more precise figure for the dose evaluation. In addition, a p lateau dose is defined, and these two dose descriptors are discussed a nd compared.