FREQUENCY-RESPONSE MEASUREMENT OF DIGITAL-COMMUNICATIONS ANALYZER PLUG-IN MODULES

Authors
Citation
R. Park et Pd. Hale, FREQUENCY-RESPONSE MEASUREMENT OF DIGITAL-COMMUNICATIONS ANALYZER PLUG-IN MODULES, HEWLETT-PAC, 47(6), 1996, pp. 37-40
Citations number
2
Categorie Soggetti
Engineering, Eletrical & Electronic
Journal title
HEWLETT-PACKARD JOURNAL
ISSN journal
00181153 → ACNP
Volume
47
Issue
6
Year of publication
1996
Pages
37 - 40
Database
ISI
SICI code
0018-1153(1996)47:6<37:FMODAP>2.0.ZU;2-Y
Abstract
it has been extremely difficult to characterize the SONET/SDH standard receiver with tolerances of +/- 0.3 dB. This paper describes a method for calibrating photoreceiver frequency response with the low inheren t uncertainty of the U.S. National Institute of Standards and Technolo gy Nd:YAG heterodyne system and transferring this calibration to a pro duction test system while maintaining a low uncertainty.