FAST MULTICHANNEL STABILIZATION OF THE MAGNETIC-RESONANCE IN A MAGNETO-RESONANCE SPECTROMETER

Citation
Yv. Borisov et al., FAST MULTICHANNEL STABILIZATION OF THE MAGNETIC-RESONANCE IN A MAGNETO-RESONANCE SPECTROMETER, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 357(1), 1995, pp. 115-119
Citations number
14
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
357
Issue
1
Year of publication
1995
Pages
115 - 119
Database
ISI
SICI code
0168-9002(1995)357:1<115:FMSOTM>2.0.ZU;2-3
Abstract
A method of fast stabilization of the magnetic resonance in a magnetic resonance spectrometer has been developed and tested. The source of t he oscillatory field is the synthesizer which frequency follows variat ions of the average magnetic field in spectrometer. To search for the neutron electric dipole moment in the experimental setup under unfavor able conditions in the reactor experimental hall a stability of the ma gnetic resonance has been achieved that is equivalent to a magnetic fi eld stability at the level of approximate to 2.5 x 10(-12) T per 6 min in a volume of about 50 l with an electric field of 1.5 MV/m.