M. Runsio et al., LEFT-VENTRICULAR FUNCTION AFTER REPEATED EPISODES OF VENTRICULAR-FIBRILLATION AND DEFIBRILLATION ASSESSED BY TRANSESOPHAGEAL ECHOCARDIOGRAPHY, European heart journal, 18(1), 1997, pp. 124-131
Background Investigators studying the effects of cardioverter-defibril
lators on left ventricular systolic function have given only minor att
ention to the diastolic effects. Objectives The purpose of this study
was to investigate the impact of repeated episodes of ventricular fibr
illation and defibrillation on systolic function and diastolic filling
of the left ventricle during non-thoracotomy implantation of a cardio
verter-defibrillator. Methods Systolic function and diastolic filling
of the left ventricle were assessed peri-operatively on a beat-by-beat
basis using a transoesophageal echo-Doppler technique in 12 patients
during greater than or equal to 4 episodes of ventricular fibrillation
and defibrillation. Systolic function was assessed from the fractiona
l area change and diastolic filling from the EIA ratio. Arterial blood
pressure and the ECG were recorded continuously. Results Blood pressu
re and heart rate did not change significantly throughout the procedur
e. The systolic function, similarly, was not significantly affected; t
he only changes were seen in the first two beats after defibrillation
when the mean fractional area increased from 0 . 2 +/- 0 . 01 to 0 . 4
+/- 0 . 02 and 0 . 3 +/- 0 . 02, respectively (P<0 . 001). Diastolic
filling was, however, impaired as reflected by a decrease in the E/A r
atio from 2 . 6 +/- 0 . 5 before to 1 . 6 +/- 0 . 4 (P<0 . 01) after r
epeated threshold tests. Conclusions While the combined ischaemic and
electrical trauma caused by repeated episodes of ventricular fibrillat
ion and defibrillation during the implantation of a cardioverter-defib
rillator did not cause any systolic dysfunction, diastolic filling was
significantly impaired.