FLOW-THROUGH BIOSENSOR FOR SEQUENTIAL DETERMINATION OF TOTAL AND PROSTATIC ACID-PHOSPHATASE-ACTIVITY

Citation
M. Sanchezcabezudo et al., FLOW-THROUGH BIOSENSOR FOR SEQUENTIAL DETERMINATION OF TOTAL AND PROSTATIC ACID-PHOSPHATASE-ACTIVITY, Sensors and actuators. B, Chemical, 23(1), 1995, pp. 9-15
Citations number
40
Categorie Soggetti
Engineering, Eletrical & Electronic","Instument & Instrumentation
ISSN journal
09254005
Volume
23
Issue
1
Year of publication
1995
Pages
9 - 15
Database
ISI
SICI code
0925-4005(1995)23:1<9:FBFSDO>2.0.ZU;2-Y
Abstract
Two continuous methods based on the use of a conventional flow-injecti on system and a flow-through sensor for the sequential determination o f total and prostatic acid phosphatase activity are reported. The bioc hemical basis of both methods is the hydrolysis of p-nitrophenyl-phosp hate catalysed by the analyte and photometric monitoring of the reacti on product (p-nitrophenol) at 405 nm. The enhancement of sensitivity a chieved when this species is retained on the support packed in the flo w cell makes the method suitable for the determination of the analyte activity in serum samples after 1:5 dilution. The excellent agreement of the results obtained by the proposed and a conventional method and also the recoveries obtained prove the usefulness of the proposed bios ensor.