PREDICTION OF THE TIME TO PEAK HEPATIC ENHANCEMENT TO OPTIMIZE CONTRAST-ENHANCED SPIRAL CT

Citation
T. Irie et al., PREDICTION OF THE TIME TO PEAK HEPATIC ENHANCEMENT TO OPTIMIZE CONTRAST-ENHANCED SPIRAL CT, Acta radiologica, 36(2), 1995, pp. 154-158
Citations number
6
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
Journal title
ISSN journal
02841851
Volume
36
Issue
2
Year of publication
1995
Pages
154 - 158
Database
ISI
SICI code
0284-1851(1995)36:2<154:POTTTP>2.0.ZU;2-D
Abstract
A technique for the prediction of the time to peak hepatic enhancement to optimize contrast-enhanced spiral CT has been developed. The proce dures are (a) to inject a small amount of contrast medium rapidly via the antecubital vein and measure aortic transit time (ATT) of the bolu s by single slice dynamic CT; (b) to inject contrast medium at a high rate: expecting peak hepatic enhancement to occur at ATT plus 8 second s after the end of injection; (c) to perform spiral scanning of the li ver using thin collimation and caudo-cranial table movement beginning at ATT after the end of contrast medium injection.