GONIOSCOPIC ABINTERNO ND-YLF LASER SCLEROSTOMY IN HUMAN CADAVER EYES

Citation
O. Oram et al., GONIOSCOPIC ABINTERNO ND-YLF LASER SCLEROSTOMY IN HUMAN CADAVER EYES, Ophthalmic surgery, 26(2), 1995, pp. 136-138
Citations number
20
Categorie Soggetti
Ophthalmology,Surgery
Journal title
ISSN journal
0022023X
Volume
26
Issue
2
Year of publication
1995
Pages
136 - 138
Database
ISI
SICI code
0022-023X(1995)26:2<136:GANLSI>2.0.ZU;2-D
Abstract
We explored the potential of the picosecond Nd:YLF laser to perform go nioscopic ab interno sclerostomy in human cadaver eyes. Full-thickness sclerostomies were created in 12 such eyes and confirmed by external scleral and histopathologic examination. Optimum parameters for succes sful completion of a sclerostomy were: a rectangular pattern of 0.3 mm by 0.3 mm, a repetition rate of 1000 pulses per second, an energy-per -pulse of 400 mJ, and a spot separation between 10 and 50 mum, with no focal offset. The Nd:YLF laser reliably created successful gonioscopi c sclerostomies with minimal adjacent thermal damage in human cadaver eyes.