K. Dorner et al., IMPROVED AUTOMATED LEUKOCYTE COUNTING AND DIFFERENTIAL IN NEWBORNS ACHIEVED BY THE HEMATOLOGY ANALYZER CELL-DYN 3500(TM), Clinical and laboratory haematology, 17(1), 1995, pp. 23-30
Automated leucocyte counts in newborns generated by the impedance prin
ciple are artificially affected by the high osmotic resistance of some
newborn RBC and possibly by the high normoblast numbers present durin
g the neonatal period. Erroneously high WBC counts may result. The hae
matology analyser CELL-DYN 3500(TM) (Abbott Diagnostika GmbH, Wiesbade
n-Delkenheim, Germany) has two different channels for the WBC count, a
n electrical resistivity (impedance) channel and a laseroptical channe
l. In combination with facultative extended lysis of resistant RBC bef
ore WBC count, this instrument is claimed to be very suitable for newb
orn blood analysis. We measured the WBC count and differential of 165
blood samples from newborns and cord blood on the CELL-DYN 3500T(TM).
Reticulocyte count and manual differential including normoblasts were
determined. Furthermore, some technical aspects of neonatal blood anal
ysis were evaluated: precision, cell stability, effect of incorrect bl
ood-anticoagulant ratio of small blood collecting tubes. The internal
decision making process of the CELL-DYN 3500(TM) selects the result ei
ther from the optical channel (identifies and excludes normoblasts) or
from the resistivity channel (eliminates resistant RBC). This instrum
ent gives a reliable and accurate WBC count and differential of neonat
al samples even in blood samples with normoblasts and lytic resistant
RBC. The result given by the CELL-DYN 3500(TM) can be confirmed by a s
ubsequent run in extended lyse mode.