Hj. Trappe et al., RESULTS AND EXPERIENCES OF CARDIOVERTER-D EFIBRILLATOR THERAPY AFTER 10 YEARS - OBSERVATIONS IN 353 PATIENTS, Zeitschrift fur Kardiologie, 84(3), 1995, pp. 222-231
Since January 1984, the cardioverter-defibrillator (ICD) has been impl
anted in 353 patients (pts). Epicardial ICD implantation was performed
in 207 pts and 146 pts underwent non-thoracotomy ICD implant. Overall
operative mortality was 3 % (12/353 pts) and was significantly higher
in pts with epicardial ICD implantation (11/207 pts, 5 %) than in pat
ients who received non-thoracotomy ICDs (1/146 pts, < 1%) (p < 0.05).
During a mean follow-up of 25 +/- 23 months, 63 pts (19 %) died: 8 pts
(0.8 % per year) from sudden arrhythmic death and 6 pts (0.8 % per ye
ar) suddenly without an underlying arrhythmia. Cardiac death was obser
ved in 36 pts (4 % per year) and 13 pts (2 % per year) died from other
causes. Appropriate ICD discharges were observed in 245 pts (72 %) wi
th a mean incidence of 22 +/- 44 IC discharges per pt. In pts with thi
rd-generation ICDs, 3,542 ventricular tachyarrhythmias (VTA) were obse
rved and terminated by primary ICD shocks in 778 VTA (22 %). Antitachy
cardia pacing was attempted in 2,764 VTA (78 %) and was successful in
2,484 VTA (90 %).