E. Bouveresse et al., MODIFIED ALGORITHM FOR STANDARDIZATION OF NEAR-INFRARED SPECTROMETRICINSTRUMENTS, Analytical chemistry, 67(8), 1995, pp. 1381-1389
The transfer of calibration models between near-infrared spectrometric
instruments with the algorithm proposed by Shenk and Westerhaus as a
standardization method is studied. This algorithm does not yield optim
al results when standardization samples are used that are not of the s
ame nature as the samples to be analyzed. In an attempt to improve the
results obtained with this algorithm when one uses such standardizati
on samples, five different modifications of the spectral intensity cor
rection step of the algorithm are investigated. The use of locally wei
ghted regression, which gives more weight to the standardization sampl
es close to the same spectral intensity range where the samples to be
predicted are located and less weight to the farthest samples, seems t
o be the most suitable one.