APPLICATION OF REDOX CYCLING ENHANCED CURRENT AT AN INTERDIGITATED ARRAY ELECTRODE FOR IRON-TRACE DETERMINATION IN ULTRAPURE SPECTRAL CARBON

Citation
D. Bustin et al., APPLICATION OF REDOX CYCLING ENHANCED CURRENT AT AN INTERDIGITATED ARRAY ELECTRODE FOR IRON-TRACE DETERMINATION IN ULTRAPURE SPECTRAL CARBON, Analytica chimica acta, 305(1-3), 1995, pp. 121-125
Citations number
14
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00032670
Volume
305
Issue
1-3
Year of publication
1995
Pages
121 - 125
Database
ISI
SICI code
0003-2670(1995)305:1-3<121:AORCEC>2.0.ZU;2-3
Abstract
Vertically arranged interdigitated array (IDA) electrodes with a conti nuous Pt-film basis and a 0.5 mu m thick insulation layer in the middl e separating the upper Pt-IDA electrodes (26 pairs of 400 mu m long fi ngers, 5 mu m wide with 5 mu m gaps) from the Pt-basis were applied fo r Fe trace analysis in ultrapure spectral carbon. The testing with spi ked samples as well as comparison of results with those obtained by in dependent graphite-furnace atomic absorption spectrometry (AAS) proved the reliability of the proposed sample preparation procedure and IDA signal generation in the range of Fe content 1 X 10(-5) to 1.5 X 10(-4 ) wt.%. The method can be considered as a less expensive alternative t o the AAS determination of Fe pollutants in this ultrapure material.