DESIGN AND PERFORMANCE EVALUATION OF A HIGH-RESOLUTION SMALL ANIMAL POSITRON TOMOGRAPH

Citation
S. Pavlopoulos et G. Tzanakos, DESIGN AND PERFORMANCE EVALUATION OF A HIGH-RESOLUTION SMALL ANIMAL POSITRON TOMOGRAPH, IEEE transactions on nuclear science, 43(6), 1996, pp. 3249-3255
Citations number
26
Categorie Soggetti
Nuclear Sciences & Tecnology","Engineering, Eletrical & Electronic
ISSN journal
00189499
Volume
43
Issue
6
Year of publication
1996
Part
2
Pages
3249 - 3255
Database
ISI
SICI code
0018-9499(1996)43:6<3249:DAPEOA>2.0.ZU;2-J
Abstract
We have used Monte Carlo simulation methods to design and evaluate a s mall animal Positron Tomograph, The tomograph has an inner diameter of 34.37 cm and a field of view of 20 cm (diameter) transaxially and 6 c m axially, The basic gamma-ray detector unit consists of a partially s egmented EGO block made of 18 x 16 crystals of dimensions 3.35 mm x 3. 30 mm x 30 mm, read out by a position sensitive photomultiplier tube w ith crossed anode wires, The design of the block was accomplished with our recently developed Monte Carlo simulation program for PET design, which also was used for the evaluation of the performance of the Tomo graph, The rate (trues, randoms, NEC) performance as well as the resol ution have been studied and compared to those of other tomographs.