S. Pavlopoulos et G. Tzanakos, DESIGN AND PERFORMANCE EVALUATION OF A HIGH-RESOLUTION SMALL ANIMAL POSITRON TOMOGRAPH, IEEE transactions on nuclear science, 43(6), 1996, pp. 3249-3255
We have used Monte Carlo simulation methods to design and evaluate a s
mall animal Positron Tomograph, The tomograph has an inner diameter of
34.37 cm and a field of view of 20 cm (diameter) transaxially and 6 c
m axially, The basic gamma-ray detector unit consists of a partially s
egmented EGO block made of 18 x 16 crystals of dimensions 3.35 mm x 3.
30 mm x 30 mm, read out by a position sensitive photomultiplier tube w
ith crossed anode wires, The design of the block was accomplished with
our recently developed Monte Carlo simulation program for PET design,
which also was used for the evaluation of the performance of the Tomo
graph, The rate (trues, randoms, NEC) performance as well as the resol
ution have been studied and compared to those of other tomographs.