INTERNATIONAL COMPARISON OF I-127(2)-STABILIZED HE-NE LASERS AT LAMBDA-APPROXIMATE-TO-633 NM USING THE 3RD AND THE 5TH HARMONIC LOCKING TECHNIQUE

Citation
P. Balling et al., INTERNATIONAL COMPARISON OF I-127(2)-STABILIZED HE-NE LASERS AT LAMBDA-APPROXIMATE-TO-633 NM USING THE 3RD AND THE 5TH HARMONIC LOCKING TECHNIQUE, IEEE transactions on instrumentation and measurement, 44(2), 1995, pp. 173-176
Citations number
11
Categorie Soggetti
Engineering, Eletrical & Electronic","Instument & Instrumentation
ISSN journal
00189456
Volume
44
Issue
2
Year of publication
1995
Pages
173 - 176
Database
ISI
SICI code
0018-9456(1995)44:2<173:ICOIHL>2.0.ZU;2-E
Abstract
In an international comparison of I-127(2)-stabilized helium-neon lase rs at lambda approximate to 633 mm, lasers from the CMI and the BIPM w ere stabilized by the third (THT) and the fifth (FHT) harmonic locking techniques. A frequency shift to the blue was measured when using the FHT technique. The influence of weak components of the rotational-vib rational transition P(33) 6-3 of the molecule (II)-I-127-I-129 On the frequency shift of components d, e, f, and g of the line R(127) 11-5 i s evaluated.