IN-VITRO MEASUREMENT OF THE LOADS USING A N INSTRUMENTED SPINAL FIXATION DEVICE

Citation
A. Rohlmann et al., IN-VITRO MEASUREMENT OF THE LOADS USING A N INSTRUMENTED SPINAL FIXATION DEVICE, Zeitschrift fur Orthopadie und Ihre Grenzgebiete, 133(2), 1995, pp. 136-140
Citations number
NO
Categorie Soggetti
Orthopedics
ISSN journal
00443220
Volume
133
Issue
2
Year of publication
1995
Pages
136 - 140
Database
ISI
SICI code
0044-3220(1995)133:2<136:IMOTLU>2.0.ZU;2-T
Abstract
Only little is known about the loads acting on internal spinal fixatio n devices. The forces and moments for different external loads were no w measured using instrumented implants. Measurements were performed on intact cadaver spines and after a corpectomy. Additionally the influe nce of the mounting accuracy on the distribution of the loads in the f ixators was investigated. It could be shown that the implants are not loaded symmetrically. After a corpectomy the loads on the implants are significantly higher than for an intact spine.