Dl. Hastings et al., THE EFFECT ON DIAGNOSTIC QUALITY OF USING DUAL-ISOTOPE IMAGING FOR KR-81(M) VENTILATION AND TC-99(M)-MAA PERFUSION LUNG-SCANNING, Nuclear medicine communications, 16(4), 1995, pp. 281-289
It is the practice in some centres to use dual isotope imaging to redu
ce imaging times in lung ventilation and perfusion studies with Kr-81(
m) gas and Tc-99(m)-macroaggregated albumin (Tc-99(m)-MAA) by simultan
eous acquisition of the two images. The resulting loss of image caused
by cross-talk between the two energy windows was investigated using t
wo phantoms, one with cold Tc-99(m) lesions of varying size and contra
st, and the other a uniform field of Kr-81(m). It was found that, unde
r scatter conditions typical of a patient study, the use of dual isoto
pe acquisition and a krypton generator of 470 MBq or greater resulted
in a perceptible loss of image quality with lesions up to 4 cm in diam
eter being missed. On an older camera system, without modern energy an
d linearity correction facilities, a lower generator activity of only
120 MBq was sufficient to cause image degradation even under very low
scatter conditions. Seventy-five patient studies were performed using
both single and dual isotope imaging with generator activities ranging
from 80 to 282 MBq. At these low generator activities, the studies di
d not demonstrate any differences between the images that would result
in a different diagnosis. We conclude that the use of dual isotope V/
Q scanning reduces the diagnostic value of the perfusion image if the
activity of the Kr-81(m) generator is too high, although at generator
activities of 300 MBq or less no loss of image quality will occur on m
odern camera systems.