POSITION RESOLUTION OF MSGCS WITH CATHODE READOUT

Citation
N. Amos et al., POSITION RESOLUTION OF MSGCS WITH CATHODE READOUT, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 384(2-3), 1997, pp. 342-350
Citations number
27
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
384
Issue
2-3
Year of publication
1997
Pages
342 - 350
Database
ISI
SICI code
0168-9002(1997)384:2-3<342:PROMWC>2.0.ZU;2-Z
Abstract
The performance of a telescope of micro-strip gas chambers (MSGC) has been studied in a beam of pions. Detectors with different anode pitch and with different substrates have been operated using several gas mix tures. The position resolutions obtained by reading out the cathodes f or the 200 mu m pitch is 42 mu m. For the 400 mu m pitch detectors the resolution is 42 mu m after correcting the centroid positions with a function derived from the data.