STANDARDIZATION OF PLATINUM RESISTANCE THERMOMETERS AT TEMPERATURE FIXED-POINTS IN THE RANGE 273.16 TO 693 K

Citation
Jk. Gupta et al., STANDARDIZATION OF PLATINUM RESISTANCE THERMOMETERS AT TEMPERATURE FIXED-POINTS IN THE RANGE 273.16 TO 693 K, Indian Journal of Pure & Applied Physics, 33(5), 1995, pp. 285-288
Citations number
NO
Categorie Soggetti
Physics
ISSN journal
00195596
Volume
33
Issue
5
Year of publication
1995
Pages
285 - 288
Database
ISI
SICI code
0019-5596(1995)33:5<285:SOPRTA>2.0.ZU;2-Q
Abstract
Standardization of some foregin make standard platinum resistance ther mometers (SPRTs) has been undertaken at the indigeneously developed an d fabricated temperature fixed points, i.e. at triple point of water ( TPW), melting point of gallium and freezing points of tin and zinc. Th e uncertainty in temperature measurement with these SPRTs has been eva luated and found to be +/-0.2 mK at TPW and +/-0.5 mK at gallium, tin and zinc points. The values of constants of SPRTs calculated from the resistance ratios at the above fixed points are well consistent with t he values given in the International Temperature Scale of 1990 (ITS-90 ). The results of these measurements show that the temperature fixed p oints established at the National Physical Laboratory, New Delhi are r eproducible within +/-0.2 mK and are comparable to the fixed points of any other standard laboratory in the world.