OPEN ACCESS ATMOSPHERIC-PRESSURE CHEMICAL-IONIZATION MASS-SPECTROMETRY FOR ROUTINE SAMPLE ANALYSIS

Citation
Lce. Taylor et al., OPEN ACCESS ATMOSPHERIC-PRESSURE CHEMICAL-IONIZATION MASS-SPECTROMETRY FOR ROUTINE SAMPLE ANALYSIS, Journal of the American Society for Mass Spectrometry, 6(5), 1995, pp. 387-393
Citations number
16
Categorie Soggetti
Chemistry Physical","Chemistry Analytical",Spectroscopy
ISSN journal
10440305
Volume
6
Issue
5
Year of publication
1995
Pages
387 - 393
Database
ISI
SICI code
1044-0305(1995)6:5<387:OAACM>2.0.ZU;2-H
Abstract
We report the introduction and use of an atmospheric pressure chemical ionization liquid chromatography-mass spectrometry instrument that ha s been designed specifically for use by the synthetic chemist on an op en access, walk-in basis. This instrument has been configured with an easy-to-use sample log-in terminal that requires the user to provide o nly a sample identification number and a user name. Sample analysis ta kes approximately 4 min and provides the synthetic and medicinal chemi st with rapid and reliable mass spectrometry analysis. Since installat ion of the system, it has analyzed an average of about 80 samples per day and has the capacity to run over 100 samples per day without the i ntervention of a specialist operator. This capability has eliminated t he need for an operator to analyze routine samples and allows the mass spectroscopist more time to deal with problem solving.