ASSAY OF THE ENANTIOMERS OF IBUTILIDE AND ARTILIDE USING SOLID-PHASE EXTRACTION, DERIVATIZATION, AND ACHIRAL CHIRAL COLUMN-SWITCHING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
Cyl. Hsu et Rr. Walters, ASSAY OF THE ENANTIOMERS OF IBUTILIDE AND ARTILIDE USING SOLID-PHASE EXTRACTION, DERIVATIZATION, AND ACHIRAL CHIRAL COLUMN-SWITCHING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY, Journal of chromatography B. Biomedical applications, 667(1), 1995, pp. 115-128
Citations number
11
Categorie Soggetti
Chemistry Analytical
Journal title
Journal of chromatography B. Biomedical applications
Ibutilide fumarate and artilide fumarate, new drugs for the treatment
of cardiac arrhythmias, each contain a stereogenic center bearing a se
condary alcohol group. Reversed-phase achiral-chiral column-switching
HPLC separations of the enantiomers of each compound were developed an
d validated for quantitation in plasma and other biofluids. The key co
mponent of the method was derivatization with 1-naphthyl isocyanate, w
hich enhanced the sensitivity of fluorescence detection and enabled th
e enantiomers to be separated on a Pirkle column (covalent 3,5-denitro
benzoyl-D-phenylglycine stationary phase). The lower limit of quantita
tion of ibutilide fumarate was typically 0.1 ng/ml or less of each ena
ntiomer in 1 ml of plasma. Two of the special features of the column-s
witching system included operation with two samples in the system at o
ne time, which reduced analysis time to 16 min/sample for ibutilide an
d 19 min/sample for artilide, and a relay-operated switching of detect
or outputs, which allowed achiral and chiral column chromatographic da
ta to be gathered from two detectors into a single data acquisition ch
annel.