R. Koncki et S. Glab, KINETIC-MODEL OF PH-BASED POTENTIOMETRIC ENZYMATIC SENSORS .4. ENZYMELOADING AND LIFETIME FACTORS, Analyst, 120(2), 1995, pp. 489-493
Two factors: enzyme loading factor and lifetime factor, are defined on
the basis of a kinetic model for pH-based enzymic sensors. The enzyme
loading factor is independent of measurement conditions, such as pH a
nd concentration of buffer or concentration of analyte. It allows a qu
antitative description of enzyme activity in the sensing membrane, and
so is helpful incomparisons of various biosensors working under diffe
rent conditions. The lifetime factor is defined in terms of the depend
ence of the enzyme loading factor with time. Both factors are universa
l parameters for pH-based enzymic sensors. The utility of these factor
s in a description of biosensors is exemplified with pH-membrane elect
rodes sensitized for urea using different methods of enzyme immobiliza
tion.