The hemopump HP 31 is an improved version of a catheter-mounted, trans
valvular, left ventricular assist device, which can be placed into the
left ventricle through the ascending aorta. The purpose of this study
was to examine the influence of hematocrit and afterload on the pump
flow. The hemopump was tested using a flow bench model filled with hep
arinized bovine blood. The measurements were performed at four various
hematocrit values: 16%, 24%, 32%, and 40%. The pump flow was measured
at each hematocrit value under increasing afterload pressures (40-120
mm Hg), by all pump speed levels (n = 7). The average pump flow at hi
ghest pump speed and lowest afterload was 5.1+/-0.3 l/min (mean +/- st
andard deviation). The influence of afterload on the pump flow was sta
tistically significant (p<0.001). The highest afterload pressure of 12
0 mm Hg caused a reduction in pump flow of 24 +/- 5%. The alterations
of hematocrit values caused no statistically significant influence on
the pump flow (p = 0.72). The results of our study enabled the constru
ction of the nomogram for the in vivo determination of the pump flow.
The in vivo performances of the hemopump can be improved through the a
fterload reduction, especially in the weaning phase of treatment. The
oxygen delivery can be improved through the increase in hematocrit val
ues without significant impairment of the pump flow.