Ajj. Debets et al., COLUMN LIQUID-CHROMATOGRAPHY COUPLED ONLINE WITH MASS-SPECTROMETRY USING POSTCOLUMN PHOSPHATE SUPPRESSION, HRC. Journal of high resolution chromatography, 18(1), 1995, pp. 45-48
The coupling of column liquid chromatography on-line with mass spectro
metry using post-column phosphate suppression is described. A membrane
suppressor is used to remove the non-volatile phosphate ions from the
mobile phase after the chromatographic separation prior to introducti
on into the mass spectrometer. Using post-column phosphate suppression
by a membrane device phosphate concentrations up to 10 mM phosphate,
at a mobile phase flow rate of 0.1 ml/min, could be removed for more t
han 99% without any decay of the sensitivity of the MS detection. Usin
g a new antidepressant tetracyclic basic compound, ORG 4428, and some
structure related compounds as model compounds electron impact as well
as chemical ionisation spectra of all compounds could be obtained. Th
e set-up could be used for several weeks running, using a phosphate co
ntaining mobile phase, without any loss of performance.