QUANTITATIVE BIOANALYSIS USING MATRIX-ASSISTED LASER-DESORPTION IONIZATION MASS-SPECTROMETRY

Citation
S. Jespersen et al., QUANTITATIVE BIOANALYSIS USING MATRIX-ASSISTED LASER-DESORPTION IONIZATION MASS-SPECTROMETRY, Journal of mass spectrometry., 30(2), 1995, pp. 357-364
Citations number
13
Categorie Soggetti
Chemistry Inorganic & Nuclear",Spectroscopy,Biophysics
ISSN journal
10765174
Volume
30
Issue
2
Year of publication
1995
Pages
357 - 364
Database
ISI
SICI code
1076-5174(1995)30:2<357:QBUMLI>2.0.ZU;2-D
Abstract
The application of matrix-assisted laser desorption/ionization mass sp ectrometry (MALDI-MS) for quantitative analysis was investigated with the use of internal standards. Three peptides/proteins in the mass ran ge 1000-12000 were tested and the effect of various internal standards was evaluated. Horse cytochrome c was used as an internal standard fo r bovine cytochrome c, melittin for renin and an undecapeptide B analo gue was employed as an internal standard for the decapeptide A. A line ar response was found between the measured peak height ratio and the a pplied amount of analyte when an appropriate internal standard was use d. The quantitative abilities of MALDI-MS were finally applied to the determination of the drug amperozide in plasma. The biological samples were prepared for analysis using liquid-liquid extraction prior to MA LDI-MS. A linear calibration graph was obtained using the C-13(4) stab le isotopically labelled amperozide as an internal standard.