S. Jespersen et al., QUANTITATIVE BIOANALYSIS USING MATRIX-ASSISTED LASER-DESORPTION IONIZATION MASS-SPECTROMETRY, Journal of mass spectrometry., 30(2), 1995, pp. 357-364
The application of matrix-assisted laser desorption/ionization mass sp
ectrometry (MALDI-MS) for quantitative analysis was investigated with
the use of internal standards. Three peptides/proteins in the mass ran
ge 1000-12000 were tested and the effect of various internal standards
was evaluated. Horse cytochrome c was used as an internal standard fo
r bovine cytochrome c, melittin for renin and an undecapeptide B analo
gue was employed as an internal standard for the decapeptide A. A line
ar response was found between the measured peak height ratio and the a
pplied amount of analyte when an appropriate internal standard was use
d. The quantitative abilities of MALDI-MS were finally applied to the
determination of the drug amperozide in plasma. The biological samples
were prepared for analysis using liquid-liquid extraction prior to MA
LDI-MS. A linear calibration graph was obtained using the C-13(4) stab
le isotopically labelled amperozide as an internal standard.