Km. Schmid et al., INTRACORONARY ULTRASOUND FOLLOWING EXCIMER-LASER ANGIOPLASTY - AN IN-VITRO STUDY IN HUMAN CORONARY-ARTERIES, European heart journal, 16(2), 1995, pp. 188-193
To study the ablation effects induced by excimer laser coronary angiop
lasty (ELCA), we examined 41 segments of nine isolated coronary arteri
es. An electronic intracoronary ultrasound device (ICUS: 20 MHz, 3.5F,
Endosonics) was positioned coaxially within the vessel. Angioplasty w
as performed using a 1.7 mm ELCA catheter (Spectranetics) which was pl
aced in the lumen and directed vertically onto the intimal surface of
the vessel (fluence: 10-50 mJ.mm-2). The laser catheter was removed af
ter each lasing cycle to allow the stepwise evaluation of the morpholo
gical effects of ELCA and to avoid reaching the adventitia. Ultrasound
images were compared with the corresponding histological specimens. I
n all cases, the ablation site could be correctly identified by ICUS.
No penetration of the adventitia was seen at histology. Ablation depth
was 0.31 +/- 0.18 mm as determined by histology and 0.34 +/- 0.18 mm
as determined by ultrasound; the diameter of the crater was 0.63 +/- 0
.21 mm, and 0.75 +/- 0.16 mm, respectively, and wall thickness was 0.6
8 +/- 0.18 mm, and 0.83 +/- 0.20 mm, respectively. A statistical relat
ionship between ultrasonic and histology measurements was only found,
however, for assessment of wall thickness (r = 0.71). Conclusions: The
identification of small ablation effects by ICUS was possible with gr
eat accuracy and ELCA could be performed without penetration of the ad
ventitial layers. However, exact quantification of the crater dimensio
ns was not possible due to limitations of the axial and lateral resolu
tion. Thus, for the guidance of ELCA by ICUS a further improvement in
the resolution capabilities of ICUS devices is mandatory.