PRECISION DENSITY METER FOR FILM DOSIMETERS

Citation
Av. Antipov et al., PRECISION DENSITY METER FOR FILM DOSIMETERS, Instruments and experimental techniques, 37(5), 1994, pp. 578-580
Citations number
8
Categorie Soggetti
Instument & Instrumentation",Engineering
ISSN journal
00204412
Volume
37
Issue
5
Year of publication
1994
Part
1
Pages
578 - 580
Database
ISI
SICI code
0020-4412(1994)37:5<578:PDMFFD>2.0.ZU;2-C
Abstract
The density meter is based on a SO PD(F)-5/150 film density meter (70 mu m thick), in which the optical density S = log(10)(I-0/I) is measur ed between similar to 10(-3) and 1.6 to estimate the radiation load on accelerator equipment (from 10(3) to 5 . 10(5) Gy. At S = 1.6 the rel ative error is about 0.2%. The pulsed source of light is a CsI(T1) sci ntillator exposed to alpha-particles from a Pu-238 source. The hardwar e and software are identical to those used in gamma spectrometers.