Y. Rozenman et al., RETRIEVAL OF OCCLUDING UNEXPANDED PALMAZ-SCHATZ STENT FROM A SAPHENOUS AORTOCORONARY VEIN GRAFT, Catheterization and cardiovascular diagnosis, 34(2), 1995, pp. 159-161
The Palmaz-Schatz coronary stent is used frequently to reduce the rate
of restenosis of balloon angioplasty in saphenous vein grafts. In man
y European centers, the stent, manually crimped on a balloon, is advan
ced across the stenosis without a protective sheath. This report descr
ibes a patient in whom an attempt to deploy a stent in the orifice of
a saphenous vein graft was complicated by dislodgement of the unexpand
ed stent from the balloon. The unexpanded stent caused immediate occlu
sion of the vein graft and severe ischemic symptoms. The stent was ret
rieved by inflation of another balloon in the graft, distal to the ste
nt, pulling the balloon toward the guiding catheter, and then withdraw
ing the whole system. (C) 1995 Wiley-Liss, Inc.