S. Tissot et al., CLINICAL VALIDATION OF THE DELTATRAC MONITORING-SYSTEM IN MECHANICALLY VENTILATED PATIENTS, Intensive care medicine, 21(2), 1995, pp. 149-153
Objective: To evaluate a monitor of pulmonary gas exchange (Deltatrac,
Datex) in a clinical setting. Design: After in vitro evaluation, comp
arison over 2 min between VO2 and VCO2 values measured by the Deltatra
c and the Douglas bag technique. Comparisons were also achieved over 8
h periods between the Deltatrac and a system using a mass-spectromete
r. Setting: Polyvalent intensive care unit (ICU 15 beds) in a 1200 bed
general hospital. Patients: Comparison with the Douglas bag technique
in 10 patients undergoing controlled ventilation. Comparison with the
mass-spectrometer system in 25 other patients undergoing controlled o
r pressure support ventilation. Measurements and results: Compared to
the results obtained by the Douglas bag technique, the bias (+/-2SD) f
or VO2 and VCO2, was -3.5 +/- 26.6 and 6.1 +/- 12.7 ml.min(-1) respect
ively. By comparison with the mass-spectrometer system, the bias for V
O2 and RQ was -5.8 +/- 16.0 ml.min and 0.018 +/- 0.048, respectively.
No drift between the two systems was observed over time. Conclusions:
The Deltatrac appears suitable for VO2 and VCO2 measurements in ventil
ated patients and equivalent to a mass-spectrometer system for long te
rm measurements.