LOOP-GAP CAVITIES FOR PULSED X-RAY ESR SPECTROMETERS

Citation
Sb. Orlinskii et al., LOOP-GAP CAVITIES FOR PULSED X-RAY ESR SPECTROMETERS, Instruments and experimental techniques, 37(5), 1994, pp. 599-600
Citations number
6
Categorie Soggetti
Instument & Instrumentation",Engineering
ISSN journal
00204412
Volume
37
Issue
5
Year of publication
1994
Part
2
Pages
599 - 600
Database
ISI
SICI code
0020-4412(1994)37:5<599:LCFPXE>2.0.ZU;2-5
Abstract
This paper describes the cavities with two loops connected via a gap f or X-ray ESR spectrometers. The cavities are tunable through similar t o 1 GHz and their coupling to a waveguide can be varied. The optimum m icrowave power for detecting echo in these cavities is about 10 dB low er that in traditional rectangular cavities.